輸尿管鏡維修等級由鏡體主體結構決定了維修等級,維修此類鏡子主要得更換鏡子的工作鏡管或者傳像束,所以通常只有大修和小修之分,而且大修概率占到了90%以上。
The maintenance level of ureteroscopy is determined by the main structure of the mirror body. Repairing this type of mirror mainly requires replacing the working mirror tube or image beam of the mirror, so there is usually only a distinction between major and minor repairs, and the probability of major repairs accounts for over 90%.
內窺鏡維修公司提醒,輸尿管鏡取石術是輸尿管鏡由尿道經膀胱進入輸尿管內,利用套石網籃或取石鉗把結石取出,或在輸尿管鏡下用氣壓彈道碎石機、激光碎石機、超聲彈道等碎石設備,在輸尿管鏡引導窺視下精確碎石,將結石擊碎后再取出。
The endoscopic maintenance company reminds that ureteroscopic lithotripsy is a procedure in which ureteroscopic lithotripsy enters the ureter from the urethra through the bladder, and the stones are taken out with a stone basket or a lithotripter, or pneumatic ballistic crusher, laser crusher, ultrasonic ballistic and other lithotripsy equipment are used under the ureteroscope, and the stones are accurately broken under the guidance of a catheter, and then taken out after being broken.
氣壓彈道碎石機、激光碎石機、超聲彈道碎石機會導致輸尿管鏡鏡頭前端的器械通道磨損崩缺,由于其使用特性必然會有在使用上的自然損耗,并且其使用壽命與操作者關系甚大。
Pneumatic ballistic crusher, laser crusher, and ultrasonic ballistic lithotripter cause wear and tear of the instrument channel at the front end of the ureteroscope lens. Due to their use characteristics, they will inevitably have natural loss in use, and their service life is closely related to the operator.
輸尿管鏡維修常見故障:
Common faults in ureteroscopy maintenance:
1、圖像情況:模糊進水、黑影、灰塵;
1. Image situation: blurry water ingress, black shadows, and dust;
2、光束:發黃斷絲、灼燒崩缺,亮度不夠;
2. Beam: Yellowing, broken wire, burning and missing, insufficient brightness;
3、鏡管:彎曲、斷裂、凹陷、脫落;
3. Mirror tube: bent, broken, dented, or detached;
4、器械通道:磨損崩缺、變形;
4. Instrument channel: worn, chipped, deformed;
5、物鏡:脫膠、磨損、脫落;
5. Objective lens: degumming, wear, detachment;
6、光錐:灼燒發黑、發黃、脫落缺失;
6. Light cone: burning, blackening, yellowing, loss of detachment;
7、鏡橋組件:脫落、卡口斷;
7. Mirror bridge component: detachment, broken bayonet;
8、目鏡保護片:破裂、崩缺、脫落缺失;
8. Eyepiece protective film: cracked, chipped, detached or missing;
9、眼罩:破裂、崩缺。
9. Eye mask: ruptured or missing.
Now that we have explained the maintenance level and common troubleshooting issues of ureteroscopy, we hope you can have a new understanding. For more information, please come to our website http://www.365shop.cc Conduct consultation to understand!