歡迎來到匠仁醫療設備有限公司網站!
高清攝像系統
High definition camera system
配備的監視器為 LED 顯示器,低功耗、低發熱量、亮度高及壽命長更能滿足臨床的應用。注意選擇顯示通道,可調節亮度、對比度、色飽和度等,菜單中 Auto Source Select 調 off,可縮短開機信號搜索時間,色溫、色彩、亮度、放大等調節菜單,一般不需要調整。
The equipped monitor is an LED display, which has low power consumption, low heat generation, high brightness, and long lifespan, making it more suitable for clinical applications. Pay attention to selecting the display channel, which can adjust brightness, contrast, color saturation, etc. In the menu, Auto Source Select can be set to off to shorten the search time for startup signals. The adjustment menu for color temperature, color, brightness, amplification, etc. generally does not need to be adjusted.
攝像系統
Camera system
為了延長攝像頭的壽命,建議使用無菌套;非高溫高壓的攝像頭不要高溫高壓;攝像插頭表面和主機插座內連接前要先處理干凈干燥;攝像頭盤繞存放,直徑大于15 cm 以上,否則可能損壞內部導線。
To extend the lifespan of the camera, it is recommended to use sterile sleeves; Do not use high temperature and high pressure for non high-temperature and high-pressure disinfection cameras; Before connecting the surface of the camera plug and the host socket, it should be cleaned and dried; The camera should be coiled and stored with a diameter greater than 15 cm, otherwise internal wires may be damaged.
冷光源
Cold light source
(1)氙燈開關的時間間隔要在20~30 min,頻繁開關會縮短氙燈的壽命。
(1) The time interval for switching xenon lamps should be between 20-30 minutes, as frequent switching can shorten the lifespan of xenon lamps.
(2)由于攝像主機是可以自動調光的,因此,在滿足手術視野亮度的情況下,盡量把光源的亮度調節旋鈕調低,這樣可以避免光纖過熱進而導致燒壞光纖和鏡子等。不要過度彎曲光纖,應盤成大圓圈存放。
(2) Due to the fact that the camera host can automatically dim, it is advisable to lower the brightness adjustment knob of the light source as much as possible while meeting the surgical field brightness requirements. This can avoid overheating of the optical fiber and damage to the optical fiber and mirror. Do not excessively bend the optical fiber, it should be stored in large circles.
(3)普通光纖可以采用氣體、液體浸泡、預真空壓力蒸氣方式進行。
(3) Ordinary optical fibers can be disinfected and sterilized using gas, liquid immersion, and pre vacuum pressure vapor methods.
(4)冷光源工作時,已連接的導光束或鏡子前端均不能把光直接照射到患者或無菌臺布等易燃物品上,有灼傷和燃燒的危險;為避免事故的發生,不用時光源亮度調待機或低亮度,長時間不用則關機。
(4) When working with a cold light source, the connected guide beam or front end of the mirror should not directly shine light on flammable items such as patients or sterile tablecloths, which poses a risk of burns and combustion; To avoid accidents, adjust the brightness of the time source to standby or low brightness when not in use. If not in use for a long time, turn off the device.
(5)非 STORZ 導光束連接本機時,要注意其插入光源部的長度,否則有可能損壞遮光板。
(5) When connecting non STORZ brand guide beams to this machine, pay attention to the length of the inserted light source part, otherwise it may damage the light shield.
常見故障分析及預防措施
Common fault analysis and preventive measures
監視器無圖像顯示
No image display on the monitor
故障原因:視頻接線錯或斷裂,顯示器電源未開,監視器通道選擇錯誤等。
Cause of malfunction: Video wiring error or breakage, monitor power not turned on, monitor channel selection error, etc.
預防措施:建議術前檢查監視器電源線、視頻連接線是否接好,打開監視器后選擇有信號的通道。
Preventive measures: It is recommended to check the power and video connection lines of the monitor before surgery, and select a channel with signal after turning on the monitor.
視野模糊、圖像中有塵粒或起霧
Blurred field of view, presence of dust particles or fog in the image
故障原因:攝像頭焦距未調好;攝像頭鏡面有臟污;鏡頭密封性不良或損壞漏水;光纖老化;光源亮度不夠;術中煙霧較多等。
Cause of malfunction: The camera focus is not adjusted properly; The camera mirror is dirty; Poor sealing of the lens or damage to water leakage; Fiber optic aging; Insufficient brightness of the light source; Excessive smoke during surgery.
預防措施:建議術前用無菌溫水浸泡或擦聚維酮碘,不要頻繁更換方法。
Preventive measures: It is recommended to soak or wipe povidone iodine with sterile warm water before surgery, and do not frequently change the disinfection and sterilization methods.
成像不清楚,有雪花、條紋或局部無圖像
Unclear imaging with snowflakes, stripes, or localized absence of images
故障原因:存在高頻干擾;攝像頭連接線損壞;攝像主機插口接觸不良;監視器故障;鏡頭內部柱狀鏡頭破碎等。
Cause of malfunction: There is high-frequency interference; Damaged camera connection cable; Poor contact of the camera host socket; Monitor malfunction; The internal cylindrical lens of the lens is broken, etc.
預防措施:建議攝像頭使用無菌套,避免攝像頭連接線與超刀、電刀連接線纏繞在一起;鏡頭需單獨存放,避免與其他器械碰撞,輕拿輕放,不能使用超聲波清洗鏡頭。
Preventive measures: It is recommended to use sterile covers for the camera to avoid entanglement of the camera connection wire with the ultra knife and electric knife connection wires; The lens needs to be stored separately to avoid collision with other equipment. Handle with care and do not use ultrasonic cleaning to clean the lens.
色彩偏差,圖像顯示偏暗
Color deviation and darker image display
故障原因:白平衡未調節;攝像頭連接線損壞;監視器色彩飽和度數據錯誤;鏡子導光性能差或光源、光纖損壞;光纖與鏡頭連接處受污未及時清理干凈等。
Cause of malfunction: White balance not adjusted; Damaged camera connection cable; Monitor color saturation data error; Poor light guiding performance of mirrors or damage to light sources and optical fibers; The connection between the optical fiber and the lens is contaminated and not cleaned up in a timely manner.
預防措施:建議接入攝像頭后調節白平衡,調整色飽和度,保證攝像頭與攝像主機連接處保持清潔。鏡頭清洗保養時,導光束接頭處需要拆卸清洗;注意調整光源亮度。
Precautions: It is recommended to adjust the white balance and color saturation after connecting the camera to ensure that the connection between the camera and the camera host is kept clean. When cleaning and maintaining the lens, the joint of the guiding beam needs to be disassembled and cleaned; Pay attention to adjusting the brightness of the light source.
鏡子前端物鏡磨損或密封性破壞,成像不清楚
The front objective of the mirror is worn or the sealing is damaged, resulting in unclear imaging
故障原因:被超刀、射頻、動力系統等輔助手術設備擊打損壞。
Cause of malfunction: Damaged by auxiliary surgical equipment such as super knife, radio frequency, and power system.
預防措施:使用輔助手術設備時,要保證在視野內操作,提高臨床醫師手術操作能力。
Preventive measures: When using auxiliary surgical equipment, it is necessary to ensure safe operation within the field of vision and improve the surgical operation ability of clinical physicians.
鏡身變形,無法成像
Mirror body deformation, unable to image
故障原因:受外力彎折或空中掉落所致。
Cause of malfunction: caused by external force bending or falling from the air.
預防措施:建議術中調整視野角度時保護好鏡子本身,不宜使蠻力操作;與鏡子接觸人員需輕拿輕放,小心保管。
Precautionary measures: It is recommended to protect the mirror itself when adjusting the visual field angle during the operation, rather than using brute force; Personnel in contact with mirrors should handle them with care and store them carefully.
樊經理:13153199508 李經理:13873135765
公司地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室 湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室